Smiths-Medical CADD-Solis Medication Safety Software 3.0

Ranked by severity, then by exploit likelihood. Click a CVE ID for its full record.

2CVEs
CVE-2016-8355
An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.
Published 2017-02-13 · Modified
9.9EPSS 0.015
CVE-2016-8358
An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. The affected software does not verify the identities at communication endpoints, which may allow a man-in-the-middle attacker to gain access to the communication channel between endpoints.
Published 2017-02-13 · Modified